APPLIED OPERATIONS/Image Acquisition
OPERATION INSTRUCTIONS
2-26
Page
Each click of the <+> or <-> button of the [PMT] LED slider increases or
decreases the PMT voltage by 5 V.
Each click of the slider section of the [PMT] LED slider varies the PMT voltage
by 25 V.
In the transmitted channel case, each click of the <+> or <-> button of the
[PMT] LED slider increases or decreases the PMT voltage by 1V.
In the transmitted channel case, each click of the slider section of the [PMT]
LED slider varies the PMT voltage by 10V.
Each click of the <+> or <-> button of the [Gain] LED slider increases or
decreases the Gain by X0.1.
Each click of the slider section of the [Gain] LED slider varies the Gain by .X1.0.
Each click of the <+> or <-> button of the [Offset] LED slider increases or
decreases Offset by 1%.
Each click of the slider section of the [Offset] LED slider varies the Offset by
5%.
Photomultiplier is used for sensor. When the light incident to
photomultiplier tube is too bright or PMT voltage is set to a high voltage,
PMT Over warning may be displayed as shown below.
This warning is displayed to protect photomultiplier tube when the light
incident to it exceeds a certain level. When it is displayed, decrease PMT
voltage setting.
NOTE
TIP