APPLIED OPERATIONS/Image Acquisition
OPERATION INSTRUCTIONS
2-52
Page
1 Setting the observation mode
1. In the [Mode] group box in the [Acquire] panel, select the [Surface] option button.
2. In the [Acquire] panel, select the XYT observation mode option button.
2 Setting the interval time
1. From the page tabs on the bottom right of the [Acquire] panel, select the [Time Series]
sub-panel. A panel as shown below will be displayed.
Fig. 2-17 [Time Series] Sub-panel
2. Set the interval time using the < > or < > button in the [Interval] text box.
An image with no interval can be acquired by entering [0] in the [Interval] text
box.
In this case, the [Interval] text box shows “Free Run” message.
Using < > button also sets the [Interval] text box to “Free Run”.
[Rest Time] text box
Shows the time after end of acquisition
of an image until the start of next
acquisition.
[N] text box
Set the number of scans using the <
>
or <
> button or by input from the
keyboard.
Under text box, it displays max. numbe
r
of slices that can be acquired with use o
f
physical memory only available at that
time.
[Interval] text box
Set the interval time using the <
> o
r
< > button or by input from the keyboard.
[Total Time] text box
Shows the total time required for image
acquisitions.