APPLIED OPERATIONS/Protocol processor
OPERATION INSTRUCTIONS
2-157
Page
Sub Mode: Sub Mode is for scan mode setting of New mode appears by click on Mode
area. A certain scan mode may not set because of hardware configuration. See
2-4-10 “Restrictions at [PAPP] window input time” for restriction of scan mode
setting.
In this case of Normal scan, there might be scanned before the scan mode
changes.
Scan Status: Change scanning position and its size with the selected scan mode. See 2-4-
10 “Restrictions at [PAPP] window input time” for restriction of Scan status
input.
Variables are available in [For] statement for X Pos, Y Pos, X Size, Y Sieze.
In addition, four rule calculation is possible.
For details about how to use variables, see One Point of 2-4-8 “2 Example of
Setting Procedure of XYT Observation for hours (Protocol using repetition
processing)”.
Examples of parameters: Default clip scan
position setting which appears when the clip
scan is selected with 512 x 512 pixel.
· X Pos : 256
· Y Pos : 256
· X Size : 256
· Y Size : 256
TIP
(0, 0) (511, 0)
(511, 511)(0, 511)
Y Pos 256
X Pos
256
X Size
256
Y Size
256