APPLIED OPERATIONS/Protocol processor
OPERATION INSTRUCTIONS
2-159
Page
One Point!
It is possible to designate folder in [Save File Name] column. For example, if
[R001 Image001] is entered, folder – [R001] is created under path of [Image File Name:]
and image file is saved under the name of [Image001].
Trigger Out: Select the trigger output from “00”, “01”, “10” and “11” in the pull-down menu.
Comment: Enter any comment as desired.
In case of the range of acceptable values are specified for a cell, it is checked according to
the specified range. If a value out of the specified range is entered, it is corrected in the
vicinity value.
Variables will be created as defined with For statement for input value. Also,
four operations are possible. Regarding the use of variables, see [One Point] in
2-4-8 Example of Setting Procedure of XYT Observation for hours (Protocol
using repetition processing).