APPLIED OPERATIONS/ Protocol processor
OPERATION INSTRUCTIONS
2-162
Page
Command Entry form Description
protocol.
Offset
Offset
ch value
Sets the Offset value of channel.
<E.g.> Enter “Offset 2, 5” to set 5 to the offset
value of channel 2.
PasteROI
PasteROI
Pastes ROI to the next track.
Note: Before the paste, Normal scan is
executed one time.
Pause
Pause
Suspends executing a protocol.
ReceiveCOM
ReceiveCOM
message
Receives RS-232C command from other PC.
The following data returns after command is
executed.
Completes normally: “OK” + Received
characters
Ends with error: “NG” + Received characters.
For further details, see 2-4-7.
Save Save Saves whole series of images.
SendCOM
SendCOM
message [ ,0/1]
Sends data through RS-232C to other PC.
The following data has to be returned in
synchronous mode,.
Completed normally: “OK” + sent characters
Ends with error: “NG” + sent characters
For further details, see 2-4-7.
Size
Size
X,
Y
Sets the image size in pixel.
<E.g.> Enter “Size 512, 512” to set 512X512 to
the image size.
Note: The images acquired before and after
executing the Size command cannot be
saved together. In the previous or
succeeding track, execute the Save,
Store, or Stop&Save command to save
the images acquired.
Note) Image size can be set at the following

Table of Contents