APPLIED OPERATIONS/ Protocol processor
OPERATION INSTRUCTIONS
2-166
Page
One Point!
It is possible to use variables in [For] statement in other cell of protocol processor.
<E.g>
It is possible to change laser intensity value every image acquiring time. When laser
intensity of argon laser is defined as variable N, it creates protocols that acquire image with
interval of step 10 by N equal to 10 to 100.
1. Enter [N = 10 To 100 Step 10] for Condition column in No.1 track.
2. Define laser intensity of argon laser as variable N for No.2 track.
3. Define return of repetition for No.3 track.
4. Select <Start> button.
Image will be acquired by changing laser intensity of argon laser by 10% as shown
below.
1st 2nd 3rd 4th 5th 10th
Laser (%) 10 20 30 40 50 100