APPLIED OPERATIONS/Protocol processor
OPERATION INSTRUCTIONS
2-173
Page
3. To stop the protocol without executing it till the end, click the <End> button.
When clicked while the protocol is executed, the <End> button performs
the same function as the <Series Done> button in the [Acquire] panel.
One Point!
The protocol can be processed using the items under the [Run] menu in the [PAPP] window.
• Start executing the protocol.
Select “Start” in the [Run] menu in the [PAPP] window.
• Suspend executing the protocol.
Select “Break” in the [Run] menu in the [PAPP] window.
• Fast forward the protocol.
Select “Skip” in the [Run] menu in the [PAPP] window.
• Stop executing the protocol.
Select “End” in the [Run] menu in the [PAPP] window.
One Point!
• When [Stop when error occurs] is selected – checked from [Open] menu, process is
stopped in case that error occurs during protocol processing at the track where the error
occurred. In case that the check is removed, the process goes to next track when error
occurred.
• When [Stop an interval remaining timer during pause] is selected – checked from [Option]
menu, the time at pause is not counted. In case that the check is removed, the time at
pause is counted and, if the time counted exceeds the time set for track, the process will
be immediately started at next track when protocol is executed again.
<End> button
TIP