APPLIED OPERATIONS/Image Processing
OPERATION INSTRUCTIONS
2-233
Page
3. Click the <DIC leveling> button.
The [Level] panel is newly created in the image window and shows an image after the
correction of the DIC level irregularities of the background.
Fig. 2-86 [Level] Panel After DIC Level Correction
4. If the DIC level irregularities cannot be reduced, repeat steps 1 to 3 above by varying
the correction factor.
The background irregularities
are solved.