Getting Started FLUOVIEW/Online Help
OPERATION INSTRUCTIONS
1-62
Page
4. While observing the images in the [Live] panel, change their pinhole diameter
settings.
Each click of the <+> or <-> button of the [C.A.] LED slider increases or
decreases the pinhole diameter by 5 nm.
Each click of the slider section of the [C.A.] LED slider varies the pinhole
diameter by 25 nm.
Each click of the <+> or <-> button of the Laser ND LED slider increases or
decreases the laser ND value by 1%.
Each click of the slider section of the Laser ND LED slider varies the laser ND
value by 5%.
To set to the Confocal Aperture value to its primal value, click the <Apply>
button in the [Dyes] sub-panel of the [Acquire] panel.
Clicking this button allows
fine adjustment of the
value.
Clicking this field allows the
value to be changed on a
large scale.
The ND value which are
usually used are displayed in
green.
Clicking the <+> or <-> button
or the field displays the set
value. The set value can be
varied by direct input from the
keyboard.
TIP
TIP