CPC 100 V1.41
2 - 44
Figure 29:
Creating a user-defined
default by customizing
the test procedure
default
Step 1 In the test card view, insert as many test cards as needed to the
currently open test procedure default.
Step 2 At the CPC 100 front panel view selector, switch to the Test Procedure
Overview.
Step 3 Press S
AVE AS DEFAULT to make this test procedure the new test
procedure default.
test card 1 test card 2 test card 3
A number of test
cards are inserted
to the currently
open test
procedure.
Switch to the Test
Procedure
Overview...
... and press
to make this test
procedure the new
test procedure
default.
Step 1
Step 2
Step 3