CPC 100 V1.41
3 - 4
Table 1:
Display of calculated
value in measurement
table
Measured
quantities
Display of calculated value in
measurement table
(m1) (m2)
V V Ratio:1, Ratio:5, Diff:
Ratio m1/m2 and phase angle
ϕm1 - ϕm2
(if phase angles are available; else "n/a"),
and differences
ΔU and Δϕ
U
.
I I Ratio:1, Ratio:5, Diff:
Ratio m1/m2 and phase angle
ϕm1 - ϕm2
(if phase angles are available; else "n/a"),
and differences
ΔI and Δϕ
I
.
V I For AC: Z or R, X
Impedance Z (magnitude in
Ω and phase
angle
ϕ in °) or R and X in Ω.
For DC: R
Resistance R (in
Ω)
else "n/a"
VI
For AC: P, Q, S
Real power P (in W) and cos
ϕ, apparent
power S (in VA) and cos
ϕ, reactive power
Q (in var) and cos
ϕ.
For DC: P
Real power P (in W)
else "n/a"