22
CONTENTS
ZW
User's Manual
Introduction
CONTENTS
Setting Option/Setting range Pages
I/O setting
[I/O]
Non
measurement
setting
[HLD.RST]
Non
measurement
setting
[HLD.RST]
Keep [KEEP]/Clamp [CLAMP]
*
p.4-28
Analog
output
[ANALOG]
Current output: MIN [MIN]
(Approx. 3.4 mA)/4 mA/5 mA/.../19
mA/20 mA/MAX [MAX]
(Approx. 21 mA) (Default: MAX)
Voltage output: MIN [MIN]
(Approx. -10.8 V)/-10 V/-9 V/.../9 V/
10 V/MAX [MAX] (Approx. 10.8 V)
(Default: MAX)
p.4-30
Binary
output
[BINARY]
MAX [MAX]
*
/MIN [MIN]
Judgment
[JUDGE]
Output object
[OUTPUT]
Output object
[OUTPUT]
TASK1 [TASK1]
*
/TASK2 [TASK2]/
TASK3 [TASK3]/TASK4 [TASK4]
p.4-18
Hysteresis
[HYS]
Hysteresis
[HYS]
0 to 99.9999 mm
(Default: 0.05 % of measurement
range)
p.4-18
Timer
[TIMER]
Timer
[TIMER]
Off
*
[OFF]/Off delay [OFF.DLY]/On
delay [ONDLY]/1 shot [1SHOT]
p.4-19
Timer time
[TIME]
1 to 5000 ms (Default: 1)
Analog output
[ANALOG]
Voltage or
current
*
[V OR C]
Voltage or
current
[V OR C]
Voltage output
*
/Current output
p.4-12
Output object
[OUTPUT]
Output object
[OUTPUT]
OFF [OFF]/TASK1 [TASK1]
*
/
TASK2 [TASK2]/TASK3 [TASK3]/
TASK4 [TASK4]
p.4-13
Focus
[FOCUS]
Focus
[FOCUS]
Off [OFF]
*
/On [ON]
p.4-14
Focus
Current/
Voltage 1
[CUR1]/
[VOLT1]
Current: 4/5/6/7/8/9/10/11/12/13/14/
15/16/17/18/19/20 mA
(Default: 4 mA)
Voltage: -10/-9/-8/-7/-6/-5/-4/-3/-2/-
1/0/1/2/3/4/5/6/7/8/9/10 V
(Default: -10 V)
Measurem
ent value 1
[MEAS1]
-999.999999 to 999.999999 mm
(Default: -0.5)
Focus
Current/
Voltage 2
[CUR2]/
[VOLT2]
Current: 4/5/6/7/8/9/10/11/12/13/14/
15/16/17/18/19/20 mA
(Default: 20 mA)
Voltage: -10/-9/-8/-7/-6/-5/-4/-3/-2/-
1/0/1/2/3/4/5/6/7/8/9/10 V
(Default: 10 V)
Measurem
ent value 2
[MEAS2]
-999.999999 to 999.999999 mm
(Default: 0.5)
Calibration
[CALIB]
Calibration
[CALIB]
Off
*
/On
p.4-16
Calibration
Current/
Voltage 1
[CUR1]/
[VOLT1]
Current: 4/5/6/7/8/9/10/11/12/13/14/
15/16/17/18/19/20 mA
(Default: 4 mA)
Voltage: -10/-9/-8/-7/-6/-5/-4/-3/-2/-
1/0/1/2/3/4/5/6/7/8/9/10 V
(Default: -10 V)
Calibration
value 1
[ADJ1]
-999 to 999 (Default: 0)
Calibration
Current/
Voltage 2
[CUR2]/
[VOLT2]
Current: 4/5/6/7/8/9/10/11/12/13/14/
15/16/17/18/19/20 mA
(Default: 20 mA)
Voltage: -10/-9/-8/-7/-6/-5/-4/-3/-2/-
1/0/1/2/3/4/5/6/7/8/9/10 V
(Default: 10 V)
Calibration
value 2
[ADJ2]
-999 to 999 (Default: 0)
*
The same output destination is set for all banks. The output destination cannot be set separately for individual banks.

Table of Contents

Other manuals for Omron ZW Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron ZW Series and is the answer not in the manual?

Omron ZW Series Specifications

General IconGeneral
BrandOmron
ModelZW Series
CategoryAccessories
LanguageEnglish

Related product manuals