262 / 374
SOCT User Manual Version 10.0 rev. A
Biometry scan programs are available in Anterior scan list.
AL – scan provides: AL, CCT, ACD, LT
ACD – scan program provides CCT, ACD
Once scan program is selected operator has to confirm type of scanned eye.
Figure 260. Type of eye selection window.
For each eye, select the type of crystalline currently present:
Phakic (natural lens) the patient has a natural crystalline lens.
Aphakic (no lens) the patient does not have any crystalline lens from birth or as a result
of surgery.
Pseudophakic eye where the patient has an IOL (intra ocular lens) substituting the crystalline.
In this case, it is very important to determine the type of material used by the surgeon and
select one of the following option
IOL PMMA
IOL Acrylate
IOL Silicon
IOL Unknown
Select the eye type
Select the type of vitreous