SOCT User Manual Version 10.0 rev. A
iTRACKING
iTracking™ technology can compensate involuntary eye movements and blinks. If iTracking is
enabled, each scan program is acquired twice and the system immediately creates an artifact-
free MC examination using the Motion Correction Technology™. The system saves 3
examinations if artifacts are detected. Two exams will be saved if the system does not detect
blinks and movements. iTracking is available for Angio OCT and 3D scans and can be enabled
or disabled by pressing the [ ] iTracking button.
After conducting an examination with iTracking, the system will display an acceptance window
with the results after Motion Correction. The windows for Angio OCT and 3D scans differ. A
reconstruction from Motion Correction exam will be shown in the window.
iTracking button separately stores status for each Angio OCT and 3D scans.
Acceptance window for a 3D scan
[Accept] On pressing [Accept] the system saves three scans. If after pressing [Accept] the MC
algorithm does not detect movement artifacts or blinks in two acquired scans, it will save the
two scans and will not generate any new MC scans.
[Rescan] Saves the 2 scans and returns to the acquisition window in order to repeat the scan.
If the second or further scans yield no satisfactory result, the user can either try again or use
the Motion Correction function based on scans from two or more iTracking attempts from
within the Results tab. See chapter 17.3 Motion correction.
[Reject] – The system will not save any scans.
[Do not ask again] When checked, it will always accept the scan for the user. This window
will not be displayed again until being re-enabled.
To enable the Acceptance window, head to Setup\Preferences\[DEVICE TYPE]
[Setup]\Parameters\ Select Checkbox Ask for acceptance window for iTracking 3D scan
under 22.5.3 Parameters tab.
Figure 29 3D Scan Acceptance Window