SOCT User Manual Version 10.0 rev. A
[TEST] – starts an automatic calibration process
After calibration the system displays a calibration summary:
If the system passes calibration, the following message is displayed in the calibration window:
Calibration successful. Next test required within 7 days”
If the system fails calibration the following message is displayed in the calibration window:
“NOTE: Topography analysis will be DISABLED due to the failure of the validation test.
Please re-run the Test. If failure persists, please contact the Service support.