SOCT User Manual Version 10.0 rev. A
measurement values for the left and right eye is significant or any problem is found
in the anterior chamber during the preliminary examination, check the correctness
of layers recognition and/or reliability indices on the check screen. If the
measurement result is not conclusive, review the inspection result by performing
measurement again or performing another inspection.
NOTE: Allow only well-trained operators to use the instrument
NOTE: When using the data taken by this instrument to select intraocular lenses,
thoroughly determine the selection by also examining cataract surgery methods
and exercising other inspections. If incorrect measurement data is used to select
intraocular lenses, further surgery might be required
NOTE: When using corneal topography or biometry data taken by this instrument
for diagnosis or determination of treatment, proceed carefully by taking
measurements multiple times and/or conducting other examinations.
NOTE: When using the data taken by this instrument for refractive correction
surgery, thoroughly determine the selection by also examining surgery methods
and exercising other inspections. Refractive correction surgery conducted according
to incorrect measurements or analysis results may result in further surgery or
severe complication such as keratectasia.
NOTE: Since simultaneous use of multiple devices can cause misdiagnosis or result
in a hazardous situation, exercise caution when using this instrument
NOTE: For Topography and Anterior scan it may be difficult to recognize the
boundaries when capturing an image of an eye with opacity or malformation such
as corneal disease, shallow anterior chamber, aphakic eye, pseudophakic eye or
dense cataract eye. In this case if required to correct the layers or reject
measurement values. Oct image
NOTE: No SOCT is intended to be used as the sole diagnostic aid in disease
identification of classification.
NOTE: The following artifact may appear on the OCT image, but this does not
indicate any failure.
NOTE: When the measurement light enters the cornea, sclera, conjunctiva or
intraocular lens perpendicularly, a bright line appears in the depth direction.
NOTE: Ghost noise may occur in areas with strong reflection such as cornea, sclera,
conjunctiva and iris.