LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Service Code Table (continued)
Code Service Code Description Corrective Action Code
1010 Display update queue error 1. Reload the device software*
2. Replace the A01 System PCB
1013 System detected unexpected UI reset 1. Replace the W18 UI Flex Cable
2. Replace the A05 UI PCB
1014 Voice prompt/audio watchdog failure 1. Reload the device software*
2. Replace the A01 System PCB
1015 USB data time-out error 1. Clear error and perform PIP
2. Replace the W18 UI Flex Cable
1c01 UI FPGA programming error Replace the A05 UI PCB
1c02 UI FPGA verification error Replace the A05 UI PCB
1c03 UI FPGA program file error 1. Reload the device software*
2. Replace the A05 UI PCB
1c04 UI ADC not functioning Replace the A05 UI PCB
A01 System PCB A05 UI PCB W18 UI Flex Cable