LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Service Code Table (continued)
(Continued on next page)
A01 System PCB
Code Service Code Description Corrective Action
3009 System could not erase oldest DM record 1. Clear memory (service mode)
2. Reload the device software*
3. Replace the A01 System PCB
300a System cannot clear DM memory 1. Reload the device software*
2. Replace the A01 System PCB
300b System error writing DM record 1. Clear memory (service mode)
2. Reload the device software*
3. Replace the A01 System PCB
300c System cannot read archived DM record 1. Clear memory (service mode)
2. Reload the device software*
3. Replace the A01 System PCB
3010 System error DM memory corrupt Replace the A01 System PCB
3fff Additional information related to error code that is
logged before 3fff
No action required; refer to error code logged
before 3fff.