LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Service Code Table (continued)
Code Service Code Description Corrective Action
3001 System cannot use data management (DM) Flash
memory
1. Reload the device software*
2. Replace the A01 System PCB
3002 System data management Flash memory corrupted 1. Clear memory (service mode)
2. Reload the device software*
3. Replace the A01 System PCB
3005 System cannot delete DM record 1. Clear memory (service mode)
2. Reload the device software*
3. Replace the A01 System PCB
3007 System cannot create new DM record 1. Clear memory (service mode)
2. Reload the device software*
3. Replace the A01 System PCB
3008 System could not store DM record 1. Clear memory (service mode)
2. Reload the device software*
3. Replace the A01 System PCB
6-17