LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Service Code Table (continued)
Code Service Code Description Corrective Action
4010 Service LED failed 1. Replace the W18 UI Flex Cable
2. Replace the A05 UI PCB
3. Replace the A04 Therapy PCB
4012 Voice Flash memory corrupted Reload the device software*
4013 Voice/font Flash memory invalid Reload the device software*
4fff Additional information related to error code that is
logged before 4fff
No action required for 4fff; refer to error code
logged before 4fff.
5003 System watchdog failed Replace the A01 System PCB
5004 System watchdog reset 1. Reload the device software*
2. Replace the A01 System PCB
5005 System CPU error during boot-up Replace the A01 System PCB
5006 System RAM failed during boot-up Replace the A01 System PCB
5007 System checksum failure during boot-up 1. Reload the device software*
2. Replace the A01 System PCB
5008 System boot program failure Replace the A01 System PCB
A01 System PCB A04 Therapy PCB A05 UI PCB W18 UI Flex Cable