LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Service Code Table (continued)
Code Service Code Description Corrective Action
5009 Real-time clock (RTC) access failed 1. Reload the device software*
2. Replace the A01 System PCB
500a Coin cell battery not detected 1. Replace the coin cell battery
2. Replace the A01 System PCB
500b Cannot use system NVRAM 1. Reload the device software*
2. Replace the A01 System PCB
500c System application start error 1. Reload the device software*
2. Replace the A01 System PCB
500d System software initialization time out 1. Reload the device software*
2. Replace the A01 System PCB
500e System application start error 1. Reload the device software*
2. Replace the A01 System PCB
5010 Configuration mismatch Replace the coin cell battery
A01 System PCB Coin Battery