LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Service Code Table (continued)
(Continued on next page)
A01 System PCB A02 PP PCB A04 Therapy PCB A05 UI PCB
Code Service Code Description Corrective Action
5032 Therapy USB disconnect 1. Clear error and perform PIP
2. Replace the A04 Therapy PCB
5033 PP USB download time out 1. Clear error and perform PIP
2. Reload the device software*
5036 USB initialization failed 1. Clear error and perform PIP
2. Reload the device software*
3. Replace the A01 System PCB
5037 USB unitization error 1. Clear error and perform PIP
2. Reload the device software*
3. Replace the A01 System PCB
5038 Cannot clear interrupt source Replace the A01 System PCB
5039 Bus error reported by USB PP channel 1. Clear error and perform PIP
2. Replace the A02 PP PCB
3. Replace the A01 System PCB
6-25