LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Service Code Table (continued)
(Continued on next page)
A01 System PCB A04 Therapy PCB
Code Service Code Description Corrective Action
8fff Additional information related to error code that is
logged before 8fff
No action required for 8fff; refer to error logged
before 8fff.
9004 Unable to initialize therapy control Replace the A01 System PCB
9005 Defib disabled Replace the A01 System PCB
9007 Shock not delivered Reload the device software*
9009 Defib charge time expired Replace the A07 Battery
900b Pacing rate out of tolerance Replace the A01 System PCB
900f Unable to initialize therapy control 1. Check the stack connector
2. Replace the A04 Therapy PCB
9011 Pacer fault 1. Reload the device software*
2. Replace the A04 Therapy PCB
9017 Pacer disabled Reload the device software*
901a Pacer rate storage corrupted Reload the device software*
6-33