LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Service Code Table (continued)
A01 System PCB A07 Battery A04 Therapy PCB
Code Service Code Description Corrective Action
800a System DSP error Reload the device software*
8013 Voice format error Reload the device software*
8014 No paddles data 1. Reload the device software*
2. Replace the A01 System PCB
801b DSP did not receive USB SOF (start of frame) interrupt 1. Clear error and perform PIP
2. Replace the A01 System PCB
8105 Impedance channel out of calibration Replace the A01 System PCB
8108 Paddles data out of sync 1. Reload the device software*
2. Replace the A01 System PCB
8109 Paddles pre-amp user test failed 1. Rerun the user test
2. Replace the A01 System PCB
810b Real impedance <-30 ohms for one second 1. Perform TCP - defibrillator calibration
2. Replace the A01 System PCB