LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Service Code Table (continued)
Code Service Code Description Corrective Action
9c29 3:00 AM redundant controls test: redundant controls
stuck on
Replace the A04 Therapy PCB
9c2a 3:00 AM redundant controls test: enable 2 stuck on Replace the A04 Therapy PCB
9c2b 3:00 AM redundant controls test: enable 1 stuck on Replace the A04 Therapy PCB
9c2c Capacitor voltage per pacing pulse too high Replace the A04 Therapy PCB
9c2d Capacitor current per pacing pulse too high Replace the A04 Therapy PCB
9c2e Cap current per pacing pulse too high Replace the A04 Therapy PCB
9c2f Pacing current and selected current out of range Replace the A04 Therapy PCB
9c30 Pacing pulse width too short Replace the A04 Therapy PCB
9c31 Pacing pulse width too long Replace the A04 Therapy PCB
9c32 Capacitor voltage and predicted capacitor voltage
mismatch
1. Energy capacitor is not connected or is open
2. Replace the A04 Therapy PCB
9c35 Therapy CPU instruction test failed Replace the A04 Therapy PCB
6-38