LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Service Code Table (continued)
Code Service Code Description Corrective Action
9c36 Therapy software stack overflow Reload the device software*
9c3a Energy capacitor overvoltage error Replace the A04 Therapy PCB
9c3b 3:00 AM redundant controls test: charge rate stuck on Replace the A04 Therapy PCB
9c3e Therapy software error Replace the A04 Therapy PCB
9c3f Therapy software error Replace the A04 Therapy PCB