LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Service Code Table (continued)
Code Service Code Description Corrective Action
a00b Printer communication lost 1. Clear error and perform PIP
2. Reload the device software*
a00e Printer initialization error Reload the device software*
b001 Invalid state request 1. Clear error and perform PIP
2. Replace A01 System PCB
b011 System behavior manager error Reload the device software*
b012 Energy cap charging time out Reload the device software*
b013 Shock advisory system error Reload the device software*
b014 Shock advisory system time out Reload the device software*
b016 Motion detect timer error Reload the device software*
b017 Shock result time out Reload the device software*
b018 USB interrupt error 1. Clear error and perform PIP.
2. Replace A01 System PCB.