LIFEPAK 20/20e Defibrillator/Monitor Replacement Procedures
Bottom Case (continued)
A13 Energy Capacitor
Removal (continued)
Note: Take care not to damage the adjacent hardware or wires when
removing the W07 Capacitor Discharge Cable.
5. Remove the cable tie retainer (229) securing the W07 Capacitor Discharge
Cable to the inductive resistor bracket (141).
6. Remove and discard the two 4-40 0.312 screws (173) from the rear of the
capacitor brackets (95).
7. Lift the capacitor (A13) out of the capacitor brackets.
8. Remove the capacitor sleeve with foam tape (162 and 93) from the
capacitor.
9. If removing the capacitor brackets (95), remove and discard the two 4-40
0.312 screws (173) from the front of the capacitor brackets (95) and remove
the capacitor brackets (95) and the capacitor shield (238) (see illustration on
next page).
A13
95
173
(x2)
Parts List Interconnect Diagram Module View Connectors View Main Assemblies
9-123