• Noise measurements (SPL, LEQ, SEL, Lden, Ltm3, Ltm5 and statistics) with Class 1 accuracy (IEC 61672-
1:2013) in the frequency range 10 Hz ÷ 20 kHz (with the SV 22 microphone)
• Two measurement ranges (Low and High) with total range 24 dBA RMS ÷ 194 dBA Peak (for the
microphone with the sensitivity 50 mV/Pa)
• Internal noise level: less than 17 dBA RMS
• Simultaneous measurements in three profiles of any channel with the independent set of IMPULSE, FAST
and SLOW detectors with standard A, C, LIN and G filters
• Digital True RMS detector with Peak detection, resolution 0.1 dB, Time Constants: SLOW, FAST,
IMPULSE
• 1/1 Octave and 1/3 Octave real time analysis (optional) - fifteen 1/1-octave filters with the centre
frequencies from 1 Hz to 16 kHz and forty-five 1/3-octave filters with the centre frequencies from 0.8 Hz to
20 kHz, Class 1 - IEC 1260
• FFT real time analysis with up to 1920 lines in 22.4 kHz band with Hanning, Rectangle, Kaiser-Bessel or
Flat Top window (optional)
• Reverberation Time analysis (RT 60) in 1/3 octave bands (optional)
• Dosimeter function (optional)
• Cross Spectrum function (optional)
• Sound Intensity function (optional)
• Wave recorder function (optional)
• Vibration measurements according to ISO 2631-1, 2 & 5 and ISO 5349-1 & 2 with Class 1 accuracy
(ISO 8041-1:2017) in the frequency range 0.5 Hz÷3 kHz (with SV 39A/L accelerometer) or 2 Hz÷ 10 kHz
(with SV 3023 M2 accelerometer)
• Two measurement ranges (Low and High) with total range 0.003 ms
-2
RMS ÷ 500 ms
-2
Peak (for the
accelerometer with the sensitivity 100 mV/g)
• Simultaneous RMS, VDV, MTVV or MAX, PEAK, P–P measurements in four channels with independent set
of filters and detector constants
• Digital True RMS & RMQ detectors with Peak detection, resolution 0.1 dB, Time Constants:
from 100 ms to 10 s
• W
d
, W
k
, W
c
, W
j
, W
m
, W
b
, W
g
(ISO 2631), W
h
(ISO 5439), HP1, HP3, HP10, Vel1, Vel3, Vel10, VelMF, Dil1,
Dil3, Dil10, KB weighting filters
• 1/1 Octave and 1/3 Octave real time analysis (optional) - fifteen 1/1-octave filters with the centre
frequencies from 1 Hz to 16 kHz and forty-five 1/3-octave filters with the centre frequencies from 0.8 Hz to
20 kHz, Class 1 - IEC 1260
• FFT real time analysis with up to 1920 lines in 22.4 kHz band with Hanning, Rectangle, Kaiser-Bessel or
Flat Top window (optional)
• RPM rotation speed measurements parallel to the vibration measurement (1 ÷ 99999) (option)
• Wave recorder function (optional)
• Ground Vibrations function (optional)
• Cross Spectrum function (optional).