EasyManua.ls Logo

Svantek SVAN 958A - APPENDIX D. DEFINITIONS AND FORMULAE OF MEASURED VALUES; D.1. SOUND LEVEL METER

Svantek SVAN 958A
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
291
SVAN 958A User Manual
Appendix D. DEFINITIONS AND FORMULAE OF MEASURED VALUES
D.1. SOUND LEVEL METER
D.1.1 Basic terms and definitions (SLM mode)
T
Current time period of the measurement in seconds.
1
T
The last second of the measurement.
e
T
Exposure time in seconds (time period during which a person is exposed to the action of
noise). This parameter can be set in the Exposure Time setup (Measurement menu). The
available values are from 1 minute to 12 hours with 1 minute step.
h8
T
Time period equal to 8 hours (28 800 seconds).
Exponential time constant in seconds for the giving time-weighting. Three time constant are
available: Slow (1000 ms), Fast (125 ms), Impulse (35 ms, but on falling values a longer time
constant of 1500 ms is applied).
W
Frequency-weighting filter (A, C, B or Z).
(
)
tp
W
Instantaneous frequency-weighted sound pressure with the weighting filter W. Sound
pressure is expressed in pascals (Pa).
(
)
tp
W
Instantaneous frequency and time-
weighted sound pressure with the
weighting filter W and time
constant
calculated from the
equation:
(
)
(
)
(
)
=
t
/t
dep
1
tp
2
WW
where:
variable of integration.
(
)
tr
Instantaneous sound pressure
depends on the
<RMS Integration> parameter:
( )
( )
( )
=
=
=
Exp nIntegratioRMS
Lin nIntegratioRMS
tp
tp
tr
W
W
0
p
Reference value (20 Pa).
( )
xlog
Represents the logarithm of x to the base 10.

Table of Contents

Related product manuals