The #9 function format is defined as follows:
#9,2,Len,<data byte> ... <data byte>
where:
Len - length of transferred file in bytes as ASCII,
<data byte> - byte of data in binary form.
Function responds with “#9,1;” on success and with “#9,0;” on failure.
A.10. CONTROL SETTING CODES
The control setting codes used in the SVAN 958 instrument (starting from the internal software version 3.6.1)
are given in the table below.
Table A.1. Control setting codes
Group name
Group
code
Code description
Unit type
U
U958 (read only)
Serial number
N
Nxxxx (read only)
Software version number * 100
W
WLxxx xxx - Meter version number * 100 (read only)
Wxxx xxx - Analyzer version number * 100 (read
only)
Microphone field correction
H
H0:n - Free field in channel n
H1:n - Diffuse field in channel n
Channel mode
Z
Z0:n - Vibration LM / Analyzer for channel n
Z1:n - Sound Level Meter / Analyzer for channel n
Calibration factor
Q
Qnnnn:c nnnn - real number with the value of the
calibration factor for channel c in dB
(-99.9 ÷ 99.9)
Measurement function
M
M1 - Level Meter
M2 - 1/1 OCTAVE analyser
M3 - 1/3 OCTAVE analyser
M4 - Sound dosimeter
M6 - FFT analyser
M8 - Reverberation Time (RT60)
M13 - FFT Cross-Spectrum
M14 - FFT Intensity
M17 - Wave recorder