EasyManua.ls Logo

Svantek SVAN 958A - E.3. DESCRIPTION OF THE DECAY CURVE RECORDING IN DIFFERENT MEASUREMENT METHODS

Svantek SVAN 958A
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
304
SVAN 958A User Manual
Fig 4. RT 30
evaluation
E.3. DESCRIPTION OF THE DECAY CURVE RECORDING IN DIFFERENT MEASUREMENT METHODS
DECAY method
This RT 60 measurement method requires omnidirectional sound source which emits pink noise in appropriate
frequency band. The most critical parameter of the omnidirectional sound source is emitted sound pressure
level as it was mentioned in the beginning of the appendix.
The graphical illustration of the data recording in this method is presented in Figure 5.
Fig 5. Data recording in the decay method of the reverberation time evaluation
t [s]
L
[dB]
t
4
t
1
L
1
L
4
5 dB
noise
margin
linear approximation
t
2
t
3
30 dB
t [s]
L
[dB]
t
e
t
s
L
2
t
trig
10 dB
L
1
registration time
t
off
t
on
L
2
L
3