4.5.4 Alarm on SMS trigger setting Alarm SMS Notification
The Alarm SMS Notific. position enables the
user to program the trigger, which generates
an alarm SMS. This position is available for
the SVAN 958A Firmware supporting modem.
The Alarm SMS Notific. position opens the
list with alarm events, defined in the
Trigger Events Setup list, that can be used
as a triggering conditions for an alarm SMS.
<ENT>
4.5.5 Alarm on e-mail trigger setting Alarm E-mail Notification
The Alarm E-mail Notif. position enables the
user to program the trigger, which generates
an alarm on e-mail. This position is available
for the SVAN 958A Firmware supporting
modem.
The Alarm E-mail Notif. position opens the
list with alarm events, defined in the
Trigger Events Setup list, that can be used
as a triggering conditions an alarm e-mail.
<ENT>
4.5.6 Definition of triggering conditions for logger and alarms Trigger Events Setup
The Trigger Events Setup position opent the
Trigger Events list, which enables the user to
define events for logger trigger and alarm
notification. These events can be defined
based on the result, calculated for any channel
and profile. Additionally it is possible to define
the Vector condition.
<ENT>
It is possible to
define two trigger
events for each
profile (vibration
input has only one
profile!).
<ENT> <ENT>
The Trigger position enables the user to switch
the trigger on and select the trigger type: Level -
, Level +, Slope -, Slope +, Grad - and Grad +.
In each interval of the measurement, defined by
Trig. Step, the triggering condition is checked
and:
if Level + is selected, the triggering condition is fulfilled only when Source has the greater value than
determined by Level, otherwise the triggering condition is not fulfilled.
if Level is selected, the triggering condition is fulfilled only when Source has the lower value than this
determined by Level, otherwise the triggering condition is not fulfilled.