EasyManua.ls Logo

Svantek SVAN 958A - Definitions and Formulas of the Results (SLM Mode)

Svantek SVAN 958A
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
292
SVAN 958A User Manual
L
T
Threshold sound level, set in the Threshold Level (path: Menu / Input / Dosimeter Setup).
The available values are as follows: None, 75dB, 80dB, 85dB or 90dB.
L
c
Criterion sound level, set in the Criterion Level (path: Menu / Input / Dosimeter Setup). The
available values are as follows: 80dB, 84dB, 85dB or 90dB.
Q
Exchange rate in decibels
equal to 2, 3, 4 or 5, set in the
Exchange Rate position (path:
Menu / Input / Dosimeter Setup).
The value of Q influences the
calculations of acoustic dose
meter results, namely DOSE,
D_8h and LAV. The exposure
rate equal to 3 complies with
ISO R 1999 “Assessment of
Occupational Noise Exposure
for Hearing Conservation
Purposes”, while Q equal to 5
complies with the American
“Occupational Safety and Health
Act” OSHA. The value of q
used in the calculations of
DOSE, D_8h and LAV is taken
from the formula:
=
=
3Qfor10
3Qfor
2log
Q
q
L(t)
Sound level (a function of time)
measured with the selected
time constant Impulse, Fast or
Slow (path: Menu / Input /
Channels Setup / Channel x
Setup) and the weighting filter (A,
C, Z or G)
0
w
p
)t(p
log20)t(L
=
L
d
(t)
Sound level (a function of time), depends on the selected threshold level.
In case when the None option
was selected
L
d
(t) = L(t)
In other cases (when the
Threshold Level is equal to
75dB, 80dB, 85dB or 90dB)
=
T
T
d
L)t(Lfor
L)t(Lfor)t(L
)t(L
D.1.2 Definitions and formulas of the results (SLM mode)
OVL
Percentage of the overloaded input signal, which occurred during the current time period of
the measurement (T)

Table of Contents

Related product manuals