117
SVAN 958A User Manual
This function returns the list of the defined (existing in the instrument) filters in the following format:
#6,type,n,name
1
, ... ,name
n
;
#6,type,W,name,v,v,...,v;
This function sets the coefficients of the new user filter named as name. The name parameter should be
unique (in the instrument there is not any other filter with the same name, otherwise it will be an error). The
function answers in the format: #6;
#6,type,R,name;
This function returns the coefficients of the user filter named as name. If the name filter does not exist, an
error occurs. The function returns in the following format: #6,type,n,v
1
,v
2
, ... ,v
n
;
#6,type,D,name;
This function deletes from the instrument the user filter named as name. If the name filter does not exist, an
error occurs. The function answers in the format: #6;
#6,type,S,name,v,v,...,v;
This function sets the user filter named as name. If the name filter already exists, its coefficients are redefined.
If the name filter does not exist, the filter is created. The function answers in the format: #6;
#6,type,C,name,first,v,v,...,v;
This function sets the coefficients in the user filter named as name starting from the first position.
If the name filter does not exist, an error occurs. The function answers in the format: #6;
#6,type,N, name
1
, name
2
;
This function changes the name of the user filter from name
1
to name
2
. The function answers
in the format: #6;
#6,type,@,chn,L;
This function returns the names of the user filters, assigned to the channel chn consecutive TOTAL values, in
the following format: #6,type,chn,3,name
1
,name
2
,name
3
;
#6,type,@,chn,pos,?;
This function returns the description record of the user filter assigned to the pos TOTAL value
of channel chn in the following format: #6,type,@,chn,pos,name,avd,cal; (the description record contains:
the name of the filter, its type and the calibration coefficient).
#6,type,@,chn,pos,*;
This function recovers the predefined filter for the pos TOTAL value of channel chn and returns the following
format: #6,type,@,chn,pos,name,avd,cal;
#6,type,@,chn,pos,name,avd,cal;
This function sets the description record of the user filter assigned to the pos TOTAL value
of channel chn in the following format: #6,type,@,chn,pos,name,avd,cal;
The returned parameters: name, avd and cal are set in the description record after the execution of the
function. In the case of an error they can differ from the current parameters of the function.