Linear operating ranges for LEQ measurements with the Free Field compensation
Two measuring ranges are available: Low and High.
The starting point at which tests of level linearity shall begin is 94.0 dB for the frequencies specifies below. For
the Low measurement range and A weighting linearity test at 31.5 Hz, the starting point is 69 dB.
Table C.1.1. Linear operating ranges for the Free Field filter and Low measurement range (for the
sinusoidal signal and microphone sensitivity 50 mV/Pa)
Table C.1.2. Linear operating ranges for the Free Field filter and High measurement range (primary) (for
the sinusoidal signal and microphone sensitivity 50 mV/Pa)
Note: For the signals with the crest factor n > 1.41 upper measuring range of the RMS
(LEQ) is reduced. The valid upper limit can be calculated according to the below given formula:
, where A is the upper limit for the sinusoidal signal
Example: For the crest factor n = 10 the upper limit is A
10
= 120 dB.
Table C.1.3. Self-generated noise for different weighting filters
*) measured with the ST 02 microphone equivalent impedance 18 pF ± 10%