274
SVAN 958A User Manual
Frequency
Compensation filter
for 0 deg incidence
angle “ SA277D
Airport
Typical compensated
response of SVAN 958A
with SA 277D for 0 deg
incidence angle
Compensated Case
Effect of SA 277D for
0 deg incidence
angle
Uncertainty
(IEC 62585:2012)
[Hz]
[dB]
[dB]
[dB]
[dB]
376
-0.06
-0.02
-0.02
0.25
387
-0.06
-0.05
-0.02
0.25
398
-0.06
-0.06
-0.01
0.25
410
-0.07
-0.07
-0.02
0.25
422
-0.07
-0.06
-0.06
0.25
434
-0.07
-0.05
-0.09
0.25
447
-0.08
-0.04
-0.06
0.25
460
-0.08
-0.02
-0.01
0.25
473
-0.09
-0.01
-0.01
0.25
487
-0.09
-0.01
-0.02
0.25
501
-0.10
-0.01
-0.01
0.25
516
-0.10
0.00
0.03
0.25
531
-0.11
0.00
0.04
0.25
546
-0.11
0.00
0.00
0.25
562
-0.12
0.01
-0.01
0.25
579
-0.13
0.01
0.04
0.25
596
-0.13
0.01
0.06
0.25
613
-0.14
0.01
0.02
0.25
631
-0.15
0.00
0.00
0.25
649
-0.15
-0.01
0.02
0.25
668
-0.16
0.00
0.05
0.25
688
-0.17
0.02
0.06
0.25
708
-0.18
0.03
0.06
0.25
729
-0.19
0.04
0.08
0.25
750
-0.20
0.05
0.08
0.25
772
-0.21
0.04
0.07
0.25
794
-0.22
0.03
0.06
0.25
818
-0.23
0.02
0.05
0.25
841
-0.24
0.03
0.07
0.25
866
-0.25
0.03
0.09
0.25
891
-0.26
0.04
0.08
0.25
917
-0.27
0.04
0.08
0.25
944
-0.28
0.04
0.05
0.25
972
-0.29
0.03
0.08
0.25
1 000
-0.30
0.01
0.05
0.25
1 029
-0.32
-0.02
-0.02
0.25
1 059
-0.33
-0.03
0.00
0.25