Connecting Communication Cables 43
2
To connect a system controller, connect the Inject Start
terminal on the autosampler as
outlined in Ta bl e 2 -2
.
Making LAC/E Connections
Use the Inject Start signal from the Waters 717plus:
If the LAC/E system contains components such as detectors that are not
connected to the IEEE-488 bus.
If the ExpertEase multi-method is set to Start by Injector.
Making Detector Connections
Waters detectors use the start of the injection for:
Auto zeroing
Start of data transfer
Detector chart marking
For specific details on programming a detector for an auto zero, refer to the detector
operators manual.
Table 2-2 System Controller Connections
Autosampler
Terminal
Controller
Controller
Terminal
Inject Start Millennium
®
Chromatography
Manager
Event In for
channel 1 or 2
Inject Start ExpertEase SAT/IN Event In for
channel 1 or 2
Inject Start ExpertEase SIM Start Inject and
COM
Inject Start 600 Series INJ and GND
Inject Start Maxima Watch Dog Event Box IN 1 or IN 2