Appendix A SCPI Command Reference
CONFigure subsystem
WCDMA Options Version 6.20
313
:CONFigure:WCDMa:MEASure:POLar:REFLevel
:CONFigure:WCDMa:MEASure:POLar:TRGLevel
:CONFigure:WCDMa:MEASure:POLar:SEGMent:NUMBer
Example
:CONF:WCDM:MEAS:RFRX:NBLO 1000
:CONF:WCDMA:MEAS:RFRX:NBLO?
Value returned in this example: 1000
.
Syntax
:CONFigure:WCDMa:MEASure:POLar:REFLevel <real1>
Parameters
real1 is a floating point real number.
The minimum value for real1
is -23.0, the maximum value is 36.0. The mini-
mum resolution for real1
is 0.1. The default value for real1 is 0.0.
Description
This command sets the reference level for the polar calibration measurement (upper
limit). The value specified for real1 is in dBm.
Query
The query form of this command will return the current setting. The string delivered
will contain one floating point real number.
Example
:CONF:WCDM:MEAS:POL:REFL -10.5
:CONF:WCDM:MEAS:POL:REFL?
Value returned in this example: -10.5
.
Syntax
:CONFigure:WCDMa:MEASure:POLar:TRGLevel <real1>
Parameters
real1 is a floating point real number.
The minimum value for real1 is -60.0, the maximum value is 36.0. The mini-
mum resolution for real1 is 0.1. The default value for real1 is 0.0
.
Description
This command sets the trigger level for the polar calibration measurement. The value
specified for real1 is in dBm.
Query
The query form of this command will return the current setting. The string delivered
will contain one floating point real number.
Example
:CONF:WCDM:MEAS:POL:TRGL -10.5
:CONF:WCDM:MEAS:POL:TRGL?
Value returned in this example: -10.5
.
Syntax
:CONFigure:WCDMa:MEASure:POLar:SEGMent:NUMBer <Int1>
Parameters
Int1 is an integer number. The minimum value is 1, the maximum value 300. The
default value is 1.
Description
This command sets the total number of segments for polar calibration measurement.
Query
The query form of this command will return the current setting of the Willtek 4400.
The string delivered will contain one integer number.
Example
:CONF:WCDM:MEAS:POL:SEGM:NUMB 1
:CONF:WCDM:MEAS:POL:SEGM:NUMB?
Value returned in this example: 1
.