SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
140
7.16.2.3
Unit and G.U.I. full demonstration (X-Ray emission
permanently disabled)
The following procedure allows a full simulation of the unit and G.U.I. functioning
without X-Ray emission (connection the PC required).
1.
Enter service menu (see chapter 8).
2.
Select the Exposition page (see paragraph 8.2).
3.
Check Disable permanently X-ray emission box.
4.
Click on the gear and save the new configuration in the EEPROM memory.
5.
Wait the unit reboot and use the G.U.I. and unit normally; the system will
perform the exam without the X-Ray emission.