Service Manual Periodic maintenance
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
214
7.
Select Pano in the Area panel.
8.
Press the button Calibrate.
9.
Each time the calibration window displays the message Waiting for an
acquisition press the X-ray button until the end of the exposure.
Note
If during calibration, the message Sensor not ready or Time out is
displayed by the Flat Panel calibration window, click on Calibrate button
and repeat the last performed calibration (Pano, ceph 1x1 or 2x2,3D 1x1 or
2x2).