EasyManua.ls Logo

Acteon X-MIND prime 3D ceph - Page 217

Acteon X-MIND prime 3D ceph
376 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Manual Periodic maintenance
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
215
Note
During the calibration, when one of the following windows is displayed,
verify that the values reported are within the tolerances:
-
Pix min: > 0.20
-
Pix max: < 0.90
and then press OK.
If they are not in the above limits, verify the copper filter positioning and
that no objects are present in the X-ray field; repeat the calibration.
10.
When the
Panoramic area calibration
is
completed
,
the message "Calibration
finished. Remove the 1.5 mm Cu filter " is displayed. This calibration will
generate the following files in the folder
C:\ProgramData\Acteon
Imaging\Panoramic X-MIND Prime Ceph
\Calibration:
-
[Sensor S/No]_2D_1x1.fmp
-
[Sensor S/No]_2D_1x1.msk
-
[Sensor S/No]_2D_1x1.Coll
-
[Sensor S/No].ini (if not present)

Table of Contents

Related product manuals