EasyManua.ls Logo

Acteon X-MIND prime 3D ceph - Page 218

Acteon X-MIND prime 3D ceph
376 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Manual Periodic maintenance
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
216
11.
Don't remove the copper filter and Select "3D" in the "Area" panel, "2x2" in the
"Resolution" panel.
12.
Press the button Calibrate.
13.
Each time the calibration window displays the message Waiting for an
acquisition press the X-ray button until the end of the exposure.
14.
When the 2x2 resolution area calibration is completed, the message
"Calibration finished. Remove the 1.5 mm Cu filter" is displayed. This calibration
will generate the following files in the folder
C:\ProgramData\Acteon
Imaging\Panoramic X-MIND Prime Ceph
\Calibration:
-
[Sensor S/No]_3D_2x2.fmp
-
[Sensor S/No]_3D_2x2.msk

Table of Contents

Related product manuals