Service Manual Periodic maintenance
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
222
43.
If the acquisitions are not OK (
X
):
-
Verify the presence of the correct calibration files (13 files named as the SN
of the sensor mounted on the unit) in the calibration folder
C:\ProgramData\Acteon Imaging\Panoramic X-MIND Prime Ceph
\Calibration
-
Verify that all the calibration options are checked in image processing
menu (see point 30 above)
-
If the acquisitions are still not OK, redo the calibration.