Service Manual Corrective maintenance
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
239
11.2.1.5
Ceph Sensor logs
These logs record the events that occur in the communication between the ceph
sensor and ASP software.
Note
Before recording the log, check that the network interface board
connected to the ceph sensor the one labelled as “Network” is configured
properly
(see paragraph 7.7.1).
To save the log:
1.
In the folder C:\Program Files (x86)\Teledyne DALSA\Sapera\Bin\ run the
program logview.exe
2.
Then try to replicate the issue and save the log file selecting in the File menu
the option save all messages.
Note
If the issue is rare open logview.exe, select the menu Options -> View GUI
settings and set 50000 in
Maximum messages shown
box.
3.
In addition in the folder C:\Windows\SysWOW64 retrieve the file logs.dat.