Service Manual – Corrective maintenance
SERVICE MANUAL • X-MIND prime 3D • (19) • 11/2019 • NXMPEN080A
11.2.8.2
MCU IP address factory reset
Follow the procedure below to restore the factory MCU IP address 192.168.0.211.
1.
Switch ON the unit.
2.
Press column UP and column DOWN keyboard buttons switch ON the unit and
keep pressed column UP and column DOWN keys until the keyboard green LED
blinks.
3.
Switch the unit OFF and then power it ON again.
4.
Set a compatible IP address on the Network interface board IP (eg.
192.168.0.16, see paragraph 7.7.1).
5.
Enter service menu (see chapter 8).
6.
Select the “Network Settings” page (see paragraph 8.1).
7.
Click on “Display default value” button and then on “Save IP address” button.
8.
Wait the unit connection.
9.
Exit the service menu by clicking on the gear.