2-94 Chapter 2
System Verification and Performance Tests
Agilent 8753ET System Verification and Performance Tests
g. Check that the trace response is 0.00 ± 0.05 dB.
h. If the trace response is out of the specified limits, repeat the “Enhanced Response
Calibration” and “Device Verification” procedures.
5. Refer to Chapter 4 , “Start Troubleshooting Here,” for more troubleshooting