EasyManua.ls Logo

Agilent Technologies 8753ET - Conversion Loss Using the Frequency Offset Mode

Agilent Technologies 8753ET
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 2-125
System Verification and Performance Tests
Agilent 8753ET System Verification and Performance Tests
Transmission Test Port Input Frequency Response from 300 kHz to 3 GHz
15.Connect the equipment as shown in Figure 2-70.
Figure 2-70 Transmission Test Port Input Frequency Response
16.Press .
17.Press .
18.Press
.
19.Press
.
20.Write the marker 1 or marker 2 value—whichever has the larger absolute
magnitude—in the performance test record.
21.If your analyzer does not have Option 006, this completes the test.
If your analyzer does have Option 006, continue with the following steps.
Power Meter Calibration for the Reflection Test Port
for 3 GHz to 6 GHz (Analyzers with Option 006)
22.Set up the analyzer as shown in Figure 2-69, using the 8481A power sensor. Cycle the
line power on the power meter. Zero and calibrate the power meter for this sensor.
23.Press .
24.Press .
25.Press . Press the appropriate softkeys
to build a calibration factor sensor table for the 8481A power sensor.
26.Press to exit the sensor calibration factor entries menu.
27.To select the 8481A power sensor, press . Press
.
Meas
INPUT PORTS
B
Sweep Setup
TRIGGER MENU
SINGLE
Marker
MODE MENU
FIXED MKR POSITION
FIXED MKR STIMULUS
300 k/m
FIXED MKR VALUE
10 x1
RETURN
REF =
FIXED MKR
Marker
MARKER 1
Marker Search
SEARCH: MAX
Marker
MARKER 2
Marker Search
SEARCH:MIN
Start 3 G/n Stop 6 G/n
Cal
PWRMTR CAL
10
x1
LOSS/SENSR LISTS
CAL FACTOR SENSOR B
DONE
USE SENSOR B
Sweep Setup
TRIGGER MENU
CONTINUOUS

Table of Contents

Other manuals for Agilent Technologies 8753ET

Related product manuals