EasyManua.ls Logo

Agilent Technologies 8753ET - Page 165

Agilent Technologies 8753ET
617 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 2-139
System Verification and Performance Tests
Agilent 8753ET System Verification and Performance Tests
Reflection and Transmission Trace Noise (Option 006 Only)
26.Connect the equipment as shown in Figure 2-78.
27.Press
.
28.Go back to step 4 on page 2-137 and repeat the rest of the procedure from this step.
In Case of Difficulty
1. Perform ADC Offset Correction Constants (Test 52)” on page 3-15.
2. Suspect the A10 digital IF assembly if both receiver channels fail.
3. Refer to Chapter 8 , “Receiver Troubleshooting,” for troubleshooting information.
Meas
REFLECTION
Power 5 x1 Avg
IF BW
3 k/m
Sweep Setup
CW FREQ
6
G/n

Table of Contents

Other manuals for Agilent Technologies 8753ET

Related product manuals