Chapter 2 2-139
System Verification and Performance Tests
Agilent 8753ET System Verification and Performance Tests
Reflection and Transmission Trace Noise (Option 006 Only)
26.Connect the equipment as shown in Figure 2-78.
27.Press
.
28.Go back to step 4 on page 2-137 and repeat the rest of the procedure from this step.
In Case of Difficulty
1. Perform ADC Offset Correction Constants (Test 52)” on page 3-15.
2. Suspect the A10 digital IF assembly if both receiver channels fail.
3. Refer to Chapter 8 , “Receiver Troubleshooting,” for troubleshooting information.
Meas
REFLECTION
Power 5 x1 Avg
IF BW
3 k/m
Sweep Setup
CW FREQ
6
G/n