2-158 Chapter 2
System Verification and Performance Tests
Agilent 8753ES Performance Test Records
Agilent Technologies Company
Model 8753ES
Serial Number
Report Number
Option(s) Date
14. Test Port Output/Input Harmonics (Analyzers with Option 002)
Test Description Specification
(dBc)
Measurement Value
(dBc)
Measurement
Uncertainty
(dB)
Test Port Output
Harmonics
2nd ≤ −25 ± 1.5
3rd ≤ −25 ± 1.5
Port 1 Input Harmonics
2nd ≤ −15 ± 1.5
3rd ≤ −30 ± 1.5
Port 2 Input Harmonics
2nd ≤ −15 ± 1.5
3rd ≤ −30 ± 1.5